Philips received a complaint on the dfm100 indicating that the device passed self-test, but it displayed cross.There was no patient involvement.The device was evaluated by customer only.Hospital engineer checked the device and did the self-test again.All tests passed but the device kept showing a cross.Customer confirmed that the display was defective.After replacing the display, device was back to work functionally.Based on the information available and the testing conducted, the cause of the reported problem was defective display.The reported problem was confirmed.The device was operational after repairs were completed.Device passed all required tests, and it remains at customer site.The investigation concludes that no further action is required at this time.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
|