Additional suspect medical device component involved in the event: product family: dbs-ipg-r-mri, upn: m365db12160, model: db-1216, serial: (b)(6), batch: 558442.A database analysis was performed for the vercise genus ipg and confirmed high impedances and an increase in the number of open circuits over time.There were no high impedances noted from the prior measurements but the database revealed 2 high impedance contacts in port l and 3 high impedance contacts in port l.High monopolar impedances may reflect a gap in connectivity indicating a possible break in the current path.The database cannot confirm the source or location of the observed high ohms or when the impedances went out of range.Program 2 possibly utilized a high impedance contact, as the log indicates more power usage during this time period when this program was used.Using programs that rely on high impedance contacts can require more power usage to maintain therapy and require more charging to avoid hibernation events.Also, therapy may not be delivered via electrode paths that measure high or out of range impedances; therefore, therapy may not be optimal when there are high impedance contact(s) used for programming.Since the ipg was not returned for analysis, the root cause of the high impedances could not be determined through the available logs.A return analysis is required to determine if device exhibits normal characteristics.Therefore, this investigation is unable to determine a probable cause for the complaint as the cause could not established.
|