Model Number DDPA2D1
Device Problems Failure to Capture (1081); Mechanical Problem (1384); Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2024
Event Type  malfunction  
Manufacturer Narrative
Continuation of d10: 694965 lead - implanted (b)(6) 2006 ; 5076-58 lead - implanted (b)(6) 2008.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that during the implant procedure the implantable cardioverter defibrillator (icd) exhibited setscrew problems as it was unable to get good contact with lead to pace.When the right ventricular (rv) lead was inserted there was no pacing occurring.  the set screw did tighten and make the clicking noise but no pacing occurred.  the device was removed and a new device was implanted. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Correction: b5, annex a product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that during the implant procedure the implantable cardioverter defibrillator (icd) exhibited setscrew problems as it was unable to get good contact with lead to pace or capture.When the right ventricular (rv) lead was inserted there was no pacing occurring.  the set screw did tighten and make the clicking noise but no pacing occurred.The device was removed and a new device was implanted. no patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBALT¿ XT DR MRI SURESCAN¿
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19056223
MDR Text Key340451665
Report Number9614453-2024-01364
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00763000178338
UDI-Public00763000178338
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDDPA2D1
Device Catalogue NumberDDPA2D1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2024
Initial Date FDA Received04/05/2024
Supplement Dates Manufacturer Received04/10/2024
Supplement Dates FDA Received05/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PJN701314V LEAD.
Patient Age67 YR
Patient SexMale
-
-