Model Number GZ-130PA
Device Problem Application Program Problem (2880)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2024
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that patients have been discharging themselves from the gz telemetry transmitter (tele) even though the key lock is on.Tech support (ts) worked with them and was not able to figure out how these were supposed to work.Sometimes, the locks will work, other times, they will not re-lock when they are unlocked unless you power cycle.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Attempt #1 03/15/2024 emailed customer via microsoft outlook for all items under the no information section.The customer responded back with complaint details, but they did not provide the patient and additional device information as requested.Central nurse's station: model: ni; sn: ni; device manufacturer date: ni; unique identifier (udi) #: ni; returned to nihon kohden: not returned.
 
Event Description
The biomedical engineer (bme) reported that patients have been discharging themselves from the gz telemetry transmitter (tele) even though the key lock is on.Tech support (ts) worked with them and was not able to figure out how these were supposed to work.Sometimes, the locks will work, other times, they will not re-lock when they are unlocked unless you power cycle.No patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GZ-130PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 116-8 560
JA  116-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key19056293
MDR Text Key340465642
Report Number8030229-2024-04207
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921117415
UDI-Public04931921117415
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153707
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGZ-130PA
Device Catalogue NumberGZ-130PA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2024
Initial Date FDA Received04/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/31/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CENTRAL NURSE'S STATION.
-
-