It was reported that during drg explant procedure on (b)(6) 2024 [related manufacturer report number: 1627487-2024-07639, 1627487-2024-07640], the right s1 lead was fractured and broken during removal.The physician attempted to remove the remaining right s1 lead and was unsuccessful.The right s1 lead is still dislodged in the foramen.No further intervention is planned at this time.It is unknown which s1 drg lead was broken and remains dislodged in the foramen.
|
Date of event is estimated.The allegation is against 1 of 2 leads; however, it is unknown which lead, therefore, all potential components are being listed.Additional components potentially involved in the event include: common device name: 50cm implant lead kit, slim tip, model: mn10450-50a, udi: (b)(4), serial: (b)(6), batch: 6367638.Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
|