Model Number 93550
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Wound Dehiscence (1154)
Event Type  Injury  
Event Description
Per the clinic, the patient experienced wound dehiscence at the magnet site (specific date not reported).Subsequently, the patient underwent revision surgery on (b)(6) 2024, in order to convert the patient to a percutaneous baha implant system.During the procedure, the internal magnet was removed under general anesthesia, and an abutment was placed on the internal fixture.
 
Manufacturer Narrative
This report is submitted on april 08, 2024.
 
Manufacturer Narrative
Correction: the correct date of explant (d6b) is (b)(6) 2024; not (b)(6) 2024, as previously reported.Per the clinic, the patient experienced a skin necrosis at the magnet site (specific date not reported).The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.This report is submitted on june 11, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIM400 IMPLANT MAGNET
Type of Device
COCHLEAR BAHA ATTRACT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
thanaletchumi manogaran
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key19057165
MDR Text Key339547315
Report Number6000034-2024-01278
Device Sequence Number1
Product Code LXB
UDI-Device Identifier09321502025294
UDI-Public(01)09321502025294(10)137712(17)230815
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131240
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number93550
Device Catalogue Number93550
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received04/07/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-