Catalog Number DYNJ50788F
Device Problems Suction Problem (2170); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2024
Event Type  malfunction  
Event Description
It was reported that a bulb irrigation syringe component broke "apart upon use.".
 
Manufacturer Narrative
It was reported that a bulb irrigation syringe component broke "apart upon use." no serious injury or adverse impact to a patient or a user was originally reported.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A sample was returned for evaluation and testing identified a lack of suction and a small base diameter which would not allow the bulb to properly connect to the syringe barrel.In an abundance of caution, and in response to an fda 483 issued for cfn (b)(4) on 22-jan-2024, this medwatch is being filed for the reported problem/issue.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
BREAST PACK
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP
three lakes drive
northfield IL 60093 2753
Manufacturer Contact
nigel vilches
three lakes drive
northfield, IL 60093-2753
MDR Report Key19057220
MDR Text Key339884366
Report Number1423395-2024-00323
Device Sequence Number1
Product Code FTN
UDI-Device Identifier40195327421039
UDI-Public40195327421039
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ50788F
Device Lot Number23HBI983
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2024
Initial Date Manufacturer Received 01/11/2024
Initial Date FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-