It was reported that a bulb irrigation syringe component broke "apart upon use." no serious injury or adverse impact to a patient or a user was originally reported.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A sample was returned for evaluation and testing identified a lack of suction and a small base diameter which would not allow the bulb to properly connect to the syringe barrel.In an abundance of caution, and in response to an fda 483 issued for cfn (b)(4) on 22-jan-2024, this medwatch is being filed for the reported problem/issue.If additional relevant information becomes available a supplemental medwatch will be filed.
|