• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOG BC; PERIPHERAL IV CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOG BC; PERIPHERAL IV CATHETER Back to Search Results
Catalog Number 382533
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2024
Event Type  malfunction  
Event Description
It was reported that bd insyte autog bc catheter sheared.The following information was provided by the initial reporter: unit manager communicated that an issue with the 20 g insyte autoguard bc wherein the actuation button which retracts needle did not function properly.The catheter was sheared.Rep asked several other clinicians if they have experienced the same issue, and they acknowledged they haven't.
 
Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Manufacturer Narrative
Investigation results: the complaint of a damaged catheter was confirmed and the cause appeared to be associated with use.The needle punctured the catheter and was protruding through the side of the catheter.What appeared to be blood residue was observed throughout the catheter tubing, catheter adapter, and needle hub.The blood residue indicates that the needle and catheter accessed the vessel.Had the observed damage existed prior to use, the catheter and needle would not have been able to access the vessel.The safety mechanism has been activated and the needle was partially retracted, but the notch in the needle was apparently caught on the puncture site in the catheter.No damage associated with the manufacturing process could be confirmed from the photographs.The complaint has been documented and will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.
 
Event Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE AUTOG BC
Type of Device
PERIPHERAL IV CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key19058768
MDR Text Key339584083
Report Number1710034-2024-00298
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903825332
UDI-Public(01)00382903825332
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number382533
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-