MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAWIRE AND WIRECLIP TORQUER; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAWIRE AND WIRECLIP TORQUER; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 3520
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2023
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 19mar2024.It was reported that the device could not cross the lesion.The 70% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending artery (lad) with a length of 60mm, blood vessel diameter was 2.5mm.A 330cm gw(5pk) flop rotawire was selected for percutaneous coronary intervention (pci).During the procedure, the device could not cross the lesion.The procedure was completed with another of the same device.No patient complications were reported.However, device analysis revealed that the spring tip section was detached.
 
Manufacturer Narrative
E1: initial reporter facility name: (b)(6) hospital ((b)(6) hospital district).Device evaluated by mfr: the device was returned for analysis.A visual and microscopic examination revealed that the spring tip section was observed detached.Additionally, tip section was observed bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE AND WIRECLIP TORQUER
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19059569
MDR Text Key339593239
Report Number2124215-2024-20920
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729195566
UDI-Public08714729195566
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3520
Device Catalogue Number3520
Device Lot Number0031351062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
Patient Weight60 KG
-
-