BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 39467-125
Device Problems Entrapment of Device (1212); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  Injury  
Event Description
It was reported that the burr became stuck on the guidewire, requiring additional hospitalization, resulting in procedure cancellation.A rotapro 1.25mm and rotawire guidewire were selected for an atherectomy procedure.The target lesion was located in the left anterior descending artery (lad).During the procedure, the rotawire guidewire could not pass through the rotapro and became stuck in the burr of the rotapro, requiring additional intervention.The issue occurred again with a second rotapro 1.50mm.The procedure was cancelled and rescheduled.The patient was admitted to the hospital beyond the standard of care.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19060071
MDR Text Key339600197
Report Number2124215-2024-20421
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729893356
UDI-Public8714729893356
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number39467-125
Device Catalogue Number39467-125
Device Lot Number0033181600
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/10/2024
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-