Catalog Number P200DNJPAMBJP
Device Problems Failure to Deliver (2338); No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Event Description
It was reported that it does not start even after switching the power on.There was no patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
B3: date of event is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
H6.Investigation codes: updated.Investigation summary: the affected device was returned for evaluation.No abnormalities in the appearance of the product related to the reported event could be confirmed.Functional test / performance test per s&r procedure.It was confirmed that it did not work even after turning on the switch, confirming the customer's complaint.The root cause is the oscillator does not operate due to leakage from the dump valve, which had come off inside the main unit.The oscillator assembly was replaced to correct the issue.P200dnamb device passed all functional and performance tests after repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC W/ALARMS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key19060427
MDR Text Key339677645
Report Number3012307300-2024-02257
Device Sequence Number1
Product Code BTL
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberP200DNJPAMBJP
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/15/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received04/08/2024
Supplement Dates Manufacturer Received04/21/2024
Supplement Dates FDA Received05/21/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse