Catalog Number P200DNJPAMBJP |
Device Problems
Failure to Deliver (2338); No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that it does not start even after switching the power on.There was no patient involvement and no patient harm/adverse event reported.
|
|
Manufacturer Narrative
|
B3: date of event is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
H6.Investigation codes: updated.Investigation summary: the affected device was returned for evaluation.No abnormalities in the appearance of the product related to the reported event could be confirmed.Functional test / performance test per s&r procedure.It was confirmed that it did not work even after turning on the switch, confirming the customer's complaint.The root cause is the oscillator does not operate due to leakage from the dump valve, which had come off inside the main unit.The oscillator assembly was replaced to correct the issue.P200dnamb device passed all functional and performance tests after repair.
|
|