• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0706-NT
Device Problems Break (1069); Crack (1135); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  malfunction  
Event Description
It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.One clip was successfully implanted.To further reduce mr, an additional clip was inserted.During the start of the deployment steps, it was observed a leak was present on the clip introducer.The physician decided to proceed with deploying the clip.The clip was successfully implanted, reducing the mr to a grade of 1-2.After the procedure, it was observed that there was a crack on the flush port of the clip introducer.When unscrewing the stopcock, part of the flush port detached from the introducer.There were no adverse patient effects and no clinically significant delay in the procedure.
 
Manufacturer Narrative
The device has been received.However, investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
All available information was investigated, and the reported cracked and broken flush port luer were confirmed via returned device analysis.The reported leak could not be replicated in a testing environment.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and the returned device analysis, the reported leak appears to be a cascading event of the reported cracked flush port.The broken flush port appears to be related to post-procedural handling as the flush port broke when unscrewing the stopcock.The investigation determined the reported cracked flush port luer to be related to a potential product quality issue.This complaint is within the scope of an exception escalation as the complaint description and device code match the specific issue described in the exception.Therefore, exception (issue) 127621 and exception (action) 132488 are referenced.The investigation evaluated the reported issue, and the engineering group determined the potential root cause to be mother nature in the form of environmental stress cracking (esc).The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19060532
MDR Text Key339788692
Report Number2135147-2024-01585
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037367
UDI-Public(01)05415067037367(17)240909(10)30911R1107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0706-NT
Device Lot Number30911R1107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received04/08/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received05/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANTED MITRACLIP; STEERABLE GUIDE CATHETER
Patient Age88 YR
Patient SexFemale
Patient Weight85 KG
-
-