MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955558
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2024
Event Type  malfunction  
Event Description
It was reported that three (3) prismax machines experienced programming errors.One (1) prismax machine delivered calcium at the incorrect rate five (5) minutes into continuous renal replacement therapy.The machine was changed back to the correct default settings to continue treatment.There was no report of patient injury associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual device was not available.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.A review of the event history log confirmed the machine worked as per configuration without any anomalies occurred during the treatment.The reported condition was verified.The cause of the condition was determined to be related to user error.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAX MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - BROOKLYN PARK
7601 northland drive
n suite 170
minneapolis MN 55428
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19067269
MDR Text Key340465798
Report Number3003504604-2024-00006
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955558
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received04/08/2024
Supplement Dates Manufacturer Received05/07/2024
Supplement Dates FDA Received05/09/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
NI.
-
-