Catalog Number 955558 |
Device Problem
Improper Flow or Infusion (2954)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/12/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that three (3) prismax machines experienced programming errors.One (1) prismax machine delivered calcium at the incorrect rate five (5) minutes into continuous renal replacement therapy.The machine was changed back to the correct default settings to continue treatment.There was no report of patient injury associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the actual device was not available.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.A review of the event history log confirmed the machine worked as per configuration without any anomalies occurred during the treatment.The reported condition was verified.The cause of the condition was determined to be related to user error.Should additional relevant information become available, a supplemental report will be submitted.
|
|