RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/18/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) medical co., ltd.(b)(6).
|
|
Event Description
|
Philips received a complaint on the v60 ventilator indicating that there was a proximal pressure sensor range error alarm.The device was reported to be in use at the time of the reported problem.No patient or user harm reported.The customer chose not to disclose any patient information from the time of the event.The customer stated that the device continued to operate after the alarm occurred and was then replaced with another device.This investigation is ongoing.
|
|
Manufacturer Narrative
|
The authorized service provider (asp) performed a run-in test, and the reported proximal pressure sensor range error alarm was not duplicated.The error code was confirmed in the event log of the device.When the device is powered off, the solenoids would have automatically zeroed.The asp performed an overall check, device cleaning, run in test, and functionality test following the inability to reproduce the device issue.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|