• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4) medical co., ltd.(b)(6).
 
Event Description
Philips received a complaint on the v60 ventilator indicating that there was a proximal pressure sensor range error alarm.The device was reported to be in use at the time of the reported problem.No patient or user harm reported.The customer chose not to disclose any patient information from the time of the event.The customer stated that the device continued to operate after the alarm occurred and was then replaced with another device.This investigation is ongoing.
 
Manufacturer Narrative
The authorized service provider (asp) performed a run-in test, and the reported proximal pressure sensor range error alarm was not duplicated.The error code was confirmed in the event log of the device.When the device is powered off, the solenoids would have automatically zeroed.The asp performed an overall check, device cleaning, run in test, and functionality test following the inability to reproduce the device issue.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19067669
MDR Text Key339993295
Report Number2518422-2024-18087
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838033832
UDI-Public00884838033832
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1076709
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received04/08/2024
Supplement Dates Manufacturer Received04/11/2024
Supplement Dates FDA Received04/26/2024
Date Device Manufactured02/25/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-