MAUDE Adverse Event Report: ABBOTT VASCULAR COPILOT® BLEEDBACK CONTROL VALVE; ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR COPILOT® BLEEDBACK CONTROL VALVE; ACCESSORIES Back to Search Results
Catalog Number 1003331
Device Problems Leak/Splash (1354); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a lesion in an unspecified artery.The copilot was attempted to be used, however, a failure with the device valve was reported resulting in blood leakage.Another copilot was attempted to be used but also could not maintain screwing of the valve and always returned back counterclockwise.Another copilot with a different lot was used to complete the procedure.There was no adverse patient effect and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The additional copilot device referenced in b5 is filed under a separate medwatch report number.
 
Event Description
It was reported that the procedure was to treat a lesion in an unspecified artery.The indeflator was attempted to be used, however, a failure with the device valve was reported resulting in blood leakage.Another indeflator with a different lot was used to complete the procedure.There was no reported adverse patient effect and there was no reported clinically significant delay in the procedure.Subsequent to the initially filed report, the following information was provided: the copilot could not maintain screwing of the valve and always returned back counterclockwise.There was no adverse patient effect and there was no clinically significant delay in the procedure.Returned device analysis identified another copilot bleedback control valve was returned.The account confirmed another copilot was attempted to be used but the same issue of leakage occurred.No additional information was provided.
 
Manufacturer Narrative
The device was returned for analysis.The reported loose/intermittent connection and the reported leak were unable to be confirmed.Device history record (dhr) and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot level product quality issue.As the reported difficulties were unable to be confirmed during return analysis, it is possible that the devices were not properly aligned and/or not fully connected/tightened while attempting to connect and/or the port of the device being connected was compromised resulting in the reported loose/intermittent connection and ultimately resulting in the reported leak; however this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported loose/intermittent connection and the reported leak.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Additionally, 19 sterile unused devices were returned (part number: 1003331, lot 60473838).13 randomly selected units were visually inspected with no anomalies noted.All 13 units were functionally tested.All connections formed a secured connection without any anomalies.The indeflator was pressurized to 400psi with no leak noted, maintained pressure.A proxy catheter was connected to the copilot with no anomalies noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COPILOT® BLEEDBACK CONTROL VALVE
Type of Device
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19068173
MDR Text Key339789982
Report Number2024168-2024-04443
Device Sequence Number1
Product Code DTL
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K991102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1003331
Device Lot Number60473838
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2024
Initial Date FDA Received04/08/2024
Supplement Dates Manufacturer Received05/06/2024
Supplement Dates FDA Received05/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-