Catalog Number AB2000 |
Device Problem
Visual Prompts will not Clear (2281)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/26/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.Component code = 4755 - aquabeam console, a reusable component of the aquabeam robotic system, controls the functionality of the high-velocity waterjet delivered by the aquabeam handpiece.
|
|
Event Description
|
A male patient underwent aquablation therapy for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation was informed that during the aquablation treatment, the aquabeam robotic system encountered an "e03 -console error" when priming a handpiece at 100% power during setup.Despite clearing the error and attempting priming multiple times with different handpieces, the error persisted.Eventually, the treatment was successfully completed by priming at 50% power and adjusting the treatment plan accordingly.The incident resulted in a surgical delay over 20 minutes.There were no adverse health consequences to the patient due to this event.
|
|
Manufacturer Narrative
|
For correction, refer to section d.4.Serial #.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|
|
Event Description
|
N/a.
|
|