ABBOTT GMBH ARCHITECT SYPHILIS TP REAGENT KIT; ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM Back to Search Results
Catalog Number 08D06-74
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2024
Event Type  malfunction  
Event Description
The customer reported false non-reactive architect syphilis tp and provided the following data for a 78 year old male patient: initial result was 0.19 (non-reactive) and repeat test was non-reactive (no value provided).Reference range: <1.00 s/co is non-reactive tppa was 1:320 (positive), trust platform generated a negative, and sysmex platform generated a positive result.There was no reported impact to patient management.
 
Manufacturer Narrative
This report is being filed on an international product, list number 08d06-74 and there is a similar product distributed in the us, list number similar us ln 8d06-31 / 41.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT SYPHILIS TP REAGENT KIT
Type of Device
ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key19075073
MDR Text Key339781594
Report Number3002809144-2024-00106
Device Sequence Number1
Product Code LIP
UDI-Device Identifier00380740161910
UDI-Public(01)00380740161910(17)240901(10)57035BE01
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K153730
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08D06-74
Device Lot Number57035BE01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/01/2024
Initial Date FDA Received04/09/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC I2000SR INST, 03M74-02, (B)(6)
Patient Age78 YR
Patient SexMale
-
-