• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-P290
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed during the device inspection, that the bronchovideoscope exhibited the light guide lens snake pipe peeled off.There were no reports of patient involvement.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, physical stress was applied to the insertion section during user handling and caused the coating to peel.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: "3.3 inspection of the endoscope 4 inspect the external surface of the entire insertion section, including the bending section and the distal end for any irregularities such as dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, and protruding objects." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19076580
MDR Text Key339793220
Report Number9610595-2024-07456
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberBF-P290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/09/2024
Supplement Dates Manufacturer Received04/15/2024
Supplement Dates FDA Received04/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-