A nurse reported that during cataract surgery with intraocular lens (iol) implantation, the iol haptic was broken.First the clinic suspected that the damage might be due to company injector and ordered new ones but thought it could also be the cartridge or a combination of them.The lens was removed and replaced with a backup lens in an initial procedure.There was no harm to the patient, only the operation time was prolonged.There are three medical device reports associated with this report.This is 1 of 3.Additional information was requested, but no further information is available.
|
Additional information provided in d.1., d.4., d.9., h.3., h.6.And h.10.The used company cartridge was returned.Inadequate viscoelastic was observed.The cartridge tip had and aneurysm on the bottom.Heavy stress was also observed.The cartridge had evidence of placement into a handpiece.The used company cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.Monarch product history records could not be reviewed because the reporter did not provide a lot number or any identification traceable to the manufacturing documentation.A qualified lens and handpiece were indicated.The viscoelastic used was not provided.It is unknown if a qualified viscoelastic was used.The root cause for the reported lens damage appears to be related to a failure to follow the instruction for use (ifu).Inadequate viscoelastic was observed in the cartridge.The cartridge tip has a large aneurysm and heavy stress.This is an indication that the lens/plunger were not in proper positions for advancement.Top coat dye stain testing was conducted with acceptable results.The ifu instructs to completely fill the cartridge with ophthalmic viscosurgical device (ovd) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact alcon.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately ½ turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The manufacturer internal reference number is: (b)(4).
|