• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. VISERA LAPARO-THORACO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. VISERA LAPARO-THORACO VIDEOSCOPE Back to Search Results
Model Number LTF-V3
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned and evaluated by olympus.In addition to the findings in b5, evaluation found the image was foggy and the focus was not obtained due to damage on the charged coupled device (ccd) unit, water tightness was lost due to a pinhole on the bending section cover, water tightness was lost due to a shaved venting connector of the light guide connector, the bending section cover adhesive was chipped, the angulation lever was deformed, the light guide lens was chipped, the video connector was cracked, the indicator on the light guide connector was not correct, and multiple parts of the scope were scratched.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although the foggy image was confirmed through evaluation of the device and the damaged objective lens was likely due to stress of repeated use, external factors, or handling, a definitive root cause of the foggy image could not be determined.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the instructions for use (ifu).If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Event Description
The customer reported the focus of the laparo-thoraco videoscope was poor.There was no reported patient harm or impact due to this event.During device evaluation at olympus, it was found the image was foggy due to a damaged objective lens.The report is being submitted due to the defect found during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA LAPARO-THORACO VIDEOSCOPE
Type of Device
LAPARO-THORACO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19078823
MDR Text Key340694746
Report Number9610595-2024-07490
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K955403
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/31/2021
Initial Date FDA Received04/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-