(b)(4).Date sent: 4/10/2024.D6a: exact implant date is unk.Assumed first day of the month.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo analysis: x-ray images of the device in vivo were reviewed by a medical safety officer.As per medical safety officer: "the images reviewed demonstrate a discontinuous linx device located below the diaphragm.".The mechanism/cause of failure is unknown as a hands-on analysis of the device is necessary to determine the cause of failure.No further investigation can be completed at this point.A manufacturing record evaluation was performed for the finished device 13076 number, and no non-conformances related to the malfunction were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what was the exact date of implant? what symptoms lead to the discovery of the discontinuous device? when did they begin? what was the date of the imaging which showed the discontinuous linx? was the device initially effective in controlling reflux? were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? did the patient undergo an mri since device implant? if so, when was the mri and what strength? did the patient have any other surgeries in the area? was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.What is the management plan? is device removal scheduled? is a replacement linx or fundoplication planned? when and if the explanation takes place can we ask that the procedure gets video recorded and the video shared? when and if the linx device is removed, may we ask that the device be returned for analysis? was ph testing performed prior to explant to confirm recurrent reflux? after implant, was the device initially effective in controlling reflux? when did the recurrent reflux begin? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|