• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. 1.5T LINX, 14B; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. 1.5T LINX, 14B; ANTI-REFLUX IMPLANT Back to Search Results
Catalog Number LXMC14
Device Problems Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Patient Problems Pyrosis/Heartburn (1883); Hernia (2240)
Event Date 03/25/2024
Event Type  Injury  
Event Description
It was reported that a patient had a linx placed by another surgeon in the (b)(6) area in (b)(6) of 2017.She had silent reflux but had a hh repair with linx for her severe lpr symptoms.These symptoms have recurred over the past 6 months or so.On an x-ray for her back problems, it was discovered that her linx is broken.
 
Manufacturer Narrative
(b)(4).Date sent: 4/10/2024.D6a: exact implant date is unk.Assumed first day of the month.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo analysis: x-ray images of the device in vivo were reviewed by a medical safety officer.As per medical safety officer: "the images reviewed demonstrate a discontinuous linx device located below the diaphragm.".The mechanism/cause of failure is unknown as a hands-on analysis of the device is necessary to determine the cause of failure.No further investigation can be completed at this point.A manufacturing record evaluation was performed for the finished device 13076 number, and no non-conformances related to the malfunction were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what was the exact date of implant? what symptoms lead to the discovery of the discontinuous device? when did they begin? what was the date of the imaging which showed the discontinuous linx? was the device initially effective in controlling reflux? were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? did the patient undergo an mri since device implant? if so, when was the mri and what strength? did the patient have any other surgeries in the area? was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.What is the management plan? is device removal scheduled? is a replacement linx or fundoplication planned? when and if the explanation takes place can we ask that the procedure gets video recorded and the video shared? when and if the linx device is removed, may we ask that the device be returned for analysis? was ph testing performed prior to explant to confirm recurrent reflux? after implant, was the device initially effective in controlling reflux? when did the recurrent reflux begin? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 5/2/2024.Additional information received: reached out to surgeon to inquire if explant has been scheduled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 14B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4545 creek road
cincinnati OH
Manufacturer Contact
kate karberg
4545 creek rd.
cincinnati, OH 45242
3035526892
MDR Report Key19078887
MDR Text Key339814299
Report Number3008766073-2024-00058
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005349
UDI-Public00855106005349
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/07/2020
Device Catalogue NumberLXMC14
Device Lot Number13076
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/27/2024
Initial Date FDA Received04/10/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Removal/Correction NumberZ-2038-2018
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-