This report is based on information provided by philips authorized service provider (asp) and has been investigated by the philips complaint handling team.Philips received a complaint on the heartstart xl indicating that the device could not be charged during self-test.The device was evaluated by the asp, according to the operational check, the battery failed and needs to be replaced.After replacing the battery, the device was returned to full functionality.Based on the information available and the testing conducted, the cause of the reported problem was a defective battery.The reported problem was confirmed.The device was operational after repairs were completed.The device passed all required tests, and it remains at the customer site.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|