Catalog Number 122140158
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Joint Dislocation (2374); Joint Laxity (4526)
Event Date 03/28/2024
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that the patient had multiple dislocations post total hip replacement.She was scheduled for conversion from a standard bearing to a dual mobility bearing to increase stability.Patient mentions to surgeon during preop that she is experiencing some groin pain also.Intraoperatively the hip is found to be relatively stable.However when the cup is checked, there appears to be some movement and is removed showing fibrous ingrowth only.The acetabulum is prepared and a new cup is placed along with a screw for adjunct fixation.A dual mobility bearing is placed and a slightly longer construct is chosen as it created a more stabile construct for the patient.There was no delay or harm to the patient.Doi: (b)(6) 2022; dor: (b)(6) 2024; affected side: right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the device lot number is unknown, therefore a¿device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX +4 10D 40IDX58OD
Type of Device
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19080183
MDR Text Key339826915
Report Number1818910-2024-07994
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295016571
UDI-Public10603295016571
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072963
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number122140158
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/29/2024
Initial Date FDA Received04/10/2024
Supplement Dates Manufacturer Received04/15/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DLT TS CER HD 12/14 40MM +1.5; PINN SECTOR W/GRIPTION 58MM
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient Weight126 KG
-
-