The baxter technician found the brakes needed to be adjusted.Per the hillrom service manual, the totalcare® bariatric bed and totalcare® bariatric plus therapy system require an effective maintenance program.We recommend that you perform semi-annual preventive maintenance (pm).Pm will minimize downtime due to excessive wear.Inspect the casters for proper mounting, excessive wear, or tread damage.Replace as needed.Verify proper operation of the brake system.Adjust as necessary.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in december 2023.It is unknown if the facility performed any other preventative maintenance on this bed.The technician adjusted the brakes to resolve the reported event.Based on this information, no further action is required.
|