ARTHROCARE CORPORATION FIRSTPASS MINI STRAIGHT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number 72290128
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2024
Event Type  malfunction  
Event Description
It was reported that during a shoulder rotator cuff repair, the surgeon had the healicoil pk 5.5 triple loaded anchor implanted at medial row using firstpass mini straight to pass the suture thru the cuff tissues after pierce thru upper jaws retrieve out from portal it broken into 2 pieces and surgeon took a lot of time locate the small piece to took it out.The procedure was completed with a surgical delay grater than 30 minutes using a back-up device.No further complications were reported.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.A visual evaluation showed that the device was not returned in any original packaging.The suture capture is missing from the upper jaw and was not returned.Bio debris is present.A functional evaluation revealed the suture passer will deploy when the trigger is initiated.A suture will not capture due to the missing suture capture.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRSTPASS MINI STRAIGHT
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19080458
MDR Text Key339896285
Report Number3006524618-2024-00139
Device Sequence Number1
Product Code NBH
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72290128
Device Lot Number2136154
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2024
Initial Date FDA Received04/10/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-