KARL STORZ SE & CO. KG ELECTRODE, BIPOLAR Back to Search Results
Model Number 011160-10
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/26/2024
Event Type  malfunction  
Event Description
It was reported that during a tur-p the electrode broke inside the patient.It was noticed that one side of the loop was not attached to the arm - the whole of the loop section was missing.The surgeon was informed immediately and checked the patient, using the scope but the broken off piece could not be found.It is assumed that the broken off part got lost in the surgeon bag.No harm to patient, user or third reported.Since the location of the broken off part is unknown, this case is deemed reportable.
 
Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
 
Manufacturer Narrative
The complained product was provided for investigation and received at the manufacturing site on the (b)(6) 2024.The evaluation revealed the following: the working electrode is broken off at the exit of the insulation.The break surfaces show a ductile appearance on the power supplying side and is molten on the passive side.The active electrode is bent towards the neutral electrode.As the passive side shows a molten area it is most likely, that the electrode got mechanically overloaded during application while the hf current has been on.This is supported by the fact that the active electrode is bent towards the neutral electrode, which also indicates the application of excessive force.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODE, BIPOLAR
Type of Device
ELECTRODE, BIPOLAR
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key19081540
MDR Text Key340574872
Report Number9610617-2024-00102
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number011160-10
Device Catalogue Number011160-10
Device Lot Number849647
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2024
Initial Date FDA Received04/10/2024
Supplement Dates Manufacturer Received04/18/2024
Supplement Dates FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-