Event Description
It was reported that during a total knee arthroplasty surgical procedure, it was observed that the robotic assisted saw interface left (sasi) device is believed to have come loose.It was reported that that connects to the saw handpiece was very loose.It was reported that during the cutting phase of procedure, the velys system had a pop up that read ¿error occurred¿.The only option was to shut down the entire robot and we were unable to complete all cuts with the robot.It was reported that there were no delays in the surgical procedure.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.(b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: investigation summary: the product was not returned to depuy synthes, however a photo was provided for review.Review of the photographic evidence revealed that the photo only shows the part and lot number and no other observation pertaining to the nature of the reported event could be identified.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.Additionally, functionality issues cannot be assessed through a photo investigation.Dhr review only covers the related service records.Service history record review result is not relevant to the current complaint and service process findings.The overall complaint was not confirmed as the photograph provided contained insufficient evidence of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SAW INTERFACE LEFT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic dr.
warsaw IN 46581
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19082336
MDR Text Key339986759
Report Number1818910-2024-08033
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10603295519461
UDI-Public10603295519461
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number451570105
Device Lot NumberJ46971
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/29/2024
Initial Date FDA Received04/10/2024
Supplement Dates Manufacturer Received05/16/2024
Supplement Dates FDA Received05/21/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/07/2022
Type of Device Usage Unknown
Patient Sequence Number1
-
-