Model Number 4471 |
Device Problems
Signal Artifact/Noise (1036); Fracture (1260); Failure to Read Input Signal (1581); Low impedance (2285); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/14/2024 |
Event Type
Injury
|
Event Description
|
It was reported that there were concerns that this right ventricular (rv) lead was fractured or dislodged.Additionally, there were erratic intrinsic amplitudes on this channel.The right ventricular autothreshold (rvat) test was suspended as a result of low amplitudes.It was noted that there were noisy signals exhibited on the rv channel and on the evoked response channel on the right atrial autothreshold (raat) test.The impedance also has slightly decreased.The impedance remains within range.Technical services (ts) suggested further evaluation of both right atrial (ra) and rv leads and a chest x-ray.The lead remains in use.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that there were concerns that this right ventricular (rv) lead was fractured or dislodged.Additionally, there were erratic intrinsic amplitudes on this channel.The right ventricular autothreshold (rvat) test was suspended as a result of low amplitudes.It was noted that there were noisy signals exhibited on the rv channel and on the evoked response channel on the right atrial autothreshold (raat) test.The impedance also has slightly decreased.The impedance remains within range.Technical services (ts) suggested further evaluation of both right atrial (ra) and rv leads and a chest x-ray.The lead remains in use.No adverse patient effects were reported.Subsequently, this lead was explanted and replaced for a left bundle branch (lbb) lead.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, visual inspection of the lead was performed.Inspection of the lead body and electrode tip found no anomalies.Laboratory testing did not identify any lead characteristics that would have resulted in the clinical observation of dislodgement.No lead issues were noted that would have made dislodgement more likely than what is described in the instructions for use as a known inherent risk of the use of this product.
|
|
Event Description
|
It was reported that there were concerns that this right ventricular (rv) lead was fractured or dislodged.Additionally, there were erratic intrinsic amplitudes on this channel.The right ventricular autothreshold (rvat) test was suspended as a result of low amplitudes.It was noted that there were noisy signals exhibited on the rv channel and on the evoked response channel on the right atrial autothreshold (raat) test.The impedance also has slightly decreased.The impedance remains within range.Technical services (ts) suggested further evaluation of both right atrial (ra) and rv leads and a chest x-ray.The lead remains in use.No adverse patient effects were reported.Subsequently, this lead was explanted and replaced for a left bundle branch (lbb) lead.No additional adverse patient effects were reported.The lead was returned for analysis.
|
|
Search Alerts/Recalls
|