Model Number 2798
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Damage to Ligament(s) (1952)
Event Type  malfunction  
Manufacturer Narrative
H3 visual findings observed one pair of restraints received for evaluation in the original packaging with ifus and a posey key.The restraints appear new or have not been put into use yet.No abnormalities have been observed.The reported issue was not confirmed.The buckles locked properly when in use with their respective straps.The cuff and bed connecting straps are found out to be within the manufacturing specifications.No abnormalities have been observed.Instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device.The ifu states to before each use, check cuff s and straps for cracks, tears, and/or excessive wear or stretch, broken buckles or locks, and/or that hook-and-loop adheres securely as these may allow patient to remove cuff.Discard if device is damaged or if unable to lock.The ifu also provided extra warning to not use this device with someone who has continued highly aggressive or combative behavior, self-destructive behavior, or deemed to be an immediate risk to others or to self.At this time there is no evidence that a manufacturing non-conformity contributed to the reported complaint.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file (b)(4).
 
Event Description
Customer reporting a complaint on product # 2799 & 2798.Customer states they have had ongoing issues with the restraints buckling after being washed.An rn got a sprained hand trying to close the buckle on a set she was using.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING TWICE-AS-TOUGH CUFFS
Type of Device
RESTRAINT, PROTECTIVE
Manufacturer (Section D)
TIDI PRODUCTS LLC
570 enterprise drive
neenah WI 54956
Manufacturer Contact
chris rahn
570 enterprise drive
neenah, WI 54956
9207514300
MDR Report Key19082621
MDR Text Key339882184
Report Number2182318-2024-00047
Device Sequence Number1
Product Code FMQ
UDI-Device Identifier10190676002709
UDI-Public10190676002709
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number2798
Device Catalogue Number2798
Device Lot Number3305T029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/14/2024
Initial Date FDA Received04/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-