Model Number L331
Device Problems Premature Discharge of Battery (1057); Over-Sensing (1438); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2024
Event Type  malfunction  
Event Description
It was reported that this implantable pacemaker was suspected to have premature battery depletion (pbd) and farfield r-wave oversensing (ffos).Furthermore, there were differing longevity estimates given over the past six months.Data analysis was performed, and there is no evidence of pbd.The device is operating and depleting normally.The power consumption is elevated due to high right ventricular (rv) and right atrial (ra) pacing by the device due to the patient condition.The device and leads remain in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI EL DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19085223
MDR Text Key340054076
Report Number2124215-2024-22011
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/24/2019
Device Model NumberL331
Device Catalogue NumberL331
Device Lot Number752687
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2024
Initial Date FDA Received04/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age90 YR
-
-