Model Number 866199 |
Device Problem
Key or Button Unresponsive/not Working (4063)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/15/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Available details indicate that the charge button is not functioning.It was determined that onsite evaluation was needed to further asses the issue.Fse evaluated the device and performed an operational test.The test failed as the charge button was not functioning.It was determined the therapy assembly board has malfunctioned.The therapy assembly board was replaced to resolve the issue.The device was returned to full functionality and remains at the customer site.
|
|
Event Description
|
Philips received a complaint on the defibrillator indicating that the charge button on the paddles was unable to be activated on the machine.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
|
|
Manufacturer Narrative
|
Reporting address line 1: (b)(6).Reporter city: (b)(6).Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone: (b)(6).Reporter phone: (b)(6).A follow up report will be submitted upon completion of the investigation.
|
|