• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC. EFFICIA DFM100; XL+DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC. EFFICIA DFM100; XL+DEFIBRILLATOR Back to Search Results
Model Number 866199
Device Problem Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2024
Event Type  malfunction  
Manufacturer Narrative
Available details indicate that the charge button is not functioning.It was determined that onsite evaluation was needed to further asses the issue.Fse evaluated the device and performed an operational test.The test failed as the charge button was not functioning.It was determined the therapy assembly board has malfunctioned.The therapy assembly board was replaced to resolve the issue.The device was returned to full functionality and remains at the customer site.
 
Event Description
Philips received a complaint on the defibrillator indicating that the charge button on the paddles was unable to be activated on the machine.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
 
Manufacturer Narrative
Reporting address line 1: (b)(6).Reporter city: (b)(6).Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone: (b)(6).Reporter phone: (b)(6).A follow up report will be submitted upon completion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFICIA DFM100
Type of Device
XL+DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH  
Manufacturer Contact
feng she
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key19085672
MDR Text Key340374951
Report Number3030677-2024-01339
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866199
Device Catalogue Number866199
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/15/2024
Initial Date FDA Received04/11/2024
Supplement Dates Manufacturer Received04/10/2024
Supplement Dates FDA Received04/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/23/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-