LIVANOVA DEUTSCHLAND GMBH HEATER COOLER 3T (230 VOLT); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Insufficient Heating (1287)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/14/2024 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that, during a procedure, the temperature in the heater-coolersystem was set to 37 and patients temperature went down.There is no report of any patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.Patient information was not provided g.5.The heater-cooler 16-02-80 is not distributed in the usa, and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in france.The 3t heater cooler was tested and no failure was found.Accorind to livanova field service engeneer the patient's rapid drop in temperature seems to come from something other than the thermal generator.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|