Model Number 16-02-80
Device Problem Insufficient Heating (1287)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2024
Event Type  malfunction  
Event Description
Livanova deutschland received a report that, during a procedure, the temperature in the heater-coolersystem was set to 37 and patients temperature went down.There is no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.Patient information was not provided g.5.The heater-cooler 16-02-80 is not distributed in the usa, and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in france.The 3t heater cooler was tested and no failure was found.Accorind to livanova field service engeneer the patient's rapid drop in temperature seems to come from something other than the thermal generator.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER COOLER 3T (230 VOLT)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key19085876
MDR Text Key339994412
Report Number9611109-2024-00177
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817901099
UDI-Public010403381790109911170830
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/15/2024
Initial Date FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-