Model Number 8888135163 |
Device Problem
Unraveled Material (1664)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
According to the reporter, when the catheter was being installed, the guide wire was damaged (coiled) as it moved forward.The catheter was not repaired.There was no leak.No cleaning was used on the device.Tego was not utilized.There was no luer adapter issue.No other defects or damages found on the product.The guide wire used was the one included in the kit.No other products being utilized with the device.Nothing else unusual observed on the device.The problem was on the guide wire only.There was no issue with the catheter itself.Flushing was not done.The catheter was replaced on the same day, and the new one worked without problems and the procedure was completed.No medical or surgical intervention was needed to prevent a permanent impairment of a function.The event did not lead to or extend patient hospitalization.There was no reported patient injury.
|
|
Manufacturer Narrative
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted one guidewire.The guidewire was unraveled.This can be caused by excessive force following the guidewire getting stuck.It was reported that the guide wire was coiled.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|