• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LLC MAHURKAR; CATHETER, HEMODIALYSIS, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LLC MAHURKAR; CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Model Number 8888135163
Device Problem Unraveled Material (1664)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
According to the reporter, when the catheter was being installed, the guide wire was damaged (coiled) as it moved forward.The catheter was not repaired.There was no leak.No cleaning was used on the device.Tego was not utilized.There was no luer adapter issue.No other defects or damages found on the product.The guide wire used was the one included in the kit.No other products being utilized with the device.Nothing else unusual observed on the device.The problem was on the guide wire only.There was no issue with the catheter itself.Flushing was not done.The catheter was replaced on the same day, and the new one worked without problems and the procedure was completed.No medical or surgical intervention was needed to prevent a permanent impairment of a function.The event did not lead to or extend patient hospitalization.There was no reported patient injury.
 
Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted one guidewire.The guidewire was unraveled.This can be caused by excessive force following the guidewire getting stuck.It was reported that the guide wire was coiled.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Type of Device
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN LLC
15 hampshire st
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN LLC
15 hampshire st
mansfield MA 02048
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19088172
MDR Text Key340682582
Report Number1282497-2024-00032
Device Sequence Number1
Product Code MPB
UDI-Device Identifier10884521006386
UDI-Public10884521006386
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K192302
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8888135163
Device Catalogue Number8888135163
Device Lot Number2226500175
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/14/2024
Initial Date FDA Received04/11/2024
Supplement Dates Manufacturer Received07/25/2024
Supplement Dates FDA Received07/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexFemale
-
-