Catalog Number 200148000 |
Device Problem
Failure to Cut (2587)
|
Patient Problem
Limb Fracture (4518)
|
Event Date 04/01/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
When using the medium and large calcar mills they seemed to have dulled in sharpness prematurely and consequently fractured the proximal femur in two total hip cases.Was surgery delayed due to the reported event? yes.If yes, number of minutes: 15.Action taken when event occurred? a cable was used to reduce the fracture.Related to (b)(4).
|
|
Manufacturer Narrative
|
Product complaint#: (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
|
|
Search Alerts/Recalls
|
|