Catalog Number 228151
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2024
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the product has not returned to depuy synthes mitek, however a photo was provided for review.¿the photo investigation revealed that truespan 12 degree peek had the plate partially deployed on the needle.No other anomalies could be observed.The overall complaint was confirmed as the observed condition of the truespan 12 degree peek would contribute to the complained device issue.Based on the investigation findings, the potential cause for the reported condition could be traced to when not inserting the needle to the proper depth for deployment which could have caused blocking insertion, when this occurred resistance may have been felt and the trigger did not pull all the way.As per ifu, it is necessary to use a calibrated probe, measure the width of the meniscal tissue to be repaired and set the adjustable depth.Also, it is important to fully squeeze the red deployment trigger while maintaining depth positioning to deliver the first implant.The implant is fully deployed when you hear an audible ¿click¿.Fully release the trigger after deployment, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H4: the date of manufacture was unknown udi: (b)(4) incomplete.
 
Event Description
It was reported by an affiliate in china that during a meniscal repair surgical procedure on (b)(6) 2024 the truespan 12 degree peek device could not fire.Another like device was used to complete the procedure.There were no adverse consequences to the patient.No additional information could be provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the product was returned to depuy synthes mitek for evaluation.The depuy synthes mitek team conducted a visual inspection and functional test of the returned device.Visual inspection found that truespan 12 degree peek had the first implant partially deployed.It does not show structural anomalies.The plates and the suture are in good conditions.The red trigger was found in a normal shape.A functional test was performed to the second plate.The applier needle was introduced into a soft tissue simulator, the red trigger was fully squeezed, the second plate was successfully deployed, no obstruction or difficulties while deploying were found.It was verified that the pusher shaft tip is sliding out completely from the applier needle.The overall complaint was confirmed as the observed condition of the truespan 12 degree peek would contribute to the complained device issue.Based on the investigation findings, the potential cause could be traced to when not inserting the needle to the proper depth for deployment which could have caused blocking insertion, when this occurred resistance may have been felt and the trigger did not pull all the way.As per ifu, it is necessary to use a calibrated probe, measure the width of the meniscal tissue to be repaired and set the adjustable depth.Also, it is important to fully squeeze the red deployment trigger while maintaining depth positioning to deliver the first implant.The implant is fully deployed when you hear an audible "click".Fully release the trigger after deployment, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4, h4: the expiration date and device manufacture date have been added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN 12 DEGREE PEEK
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key19089671
MDR Text Key340699145
Report Number1221934-2024-01192
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026012
UDI-Public10886705026012
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number228151
Device Lot Number166L363
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/07/2024
Initial Date FDA Received04/11/2024
Supplement Dates Manufacturer Received05/07/2024
05/15/2024
Supplement Dates FDA Received05/09/2024
05/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-