• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH OSTEOLOCK BONE SCREW 24MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH OSTEOLOCK BONE SCREW 24MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 5260-5-024
Device Problem Degraded (1153)
Patient Problems Pain (1994); Inadequate Osseointegration (2646); Metal Related Pathology (4530)
Event Date 12/11/2023
Event Type  Injury  
Manufacturer Narrative
An event regarding wear/metallosis involving a bone screw was reported.The event was confirmed via clinician review of the provided medical records.Method & results: -product evaluation and results: the reported device was not returned; however, photographs were provided for review.The photographs show a recently explanted metal mdm liner, poly adm liner, and 4 screws sealed inside a poly bag.Nothing remarkable can be seen on these devices in the photos.-clinician review: a review of the provided medical information by a clinical consultant indicated: "this patient underwent primary total hip arthroplasty, a revision total hip arthroplasty for polyethylene liner disassociation, with metallosis, and then another revision for acetabular loosening with metallosis.The root cause of these events cannot be determined with certainty.The causes of both acetabular loosening, polyethylene, disassociation, and metallosis, are multi factorial, including surgical technique, especially in the way preparation is made for insertion of the femoral head on the trunnion and the mdm liner into the acetabulum, as well as proper insertion of the polyethylene liner into a primary acetabular cup.Also contributing can be patient factors such as activity level and bmi, as well as implant factors." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to loosening of the acetabular component and metallosis.The reported device was not returned; however, photographs were provided for review.The photographs show a recently explanted metal mdm liner, poly adm liner, and 4 screws sealed inside a poly bag.Nothing remarkable can be seen on these devices in the photos.It was not reported what devices caused or contributed to the reported metallosis, and the stem and head in this case were competitor devices.It is possible that the metallosis generated due to motion of the loose shell against the acetabular screws or between the metal mdm liner and the shell.The event was further confirmed via clinician review of the provided medical records: "the root cause of these events cannot be determined with certainty.The causes of both acetabular loosening, polyethylene, disassociation, and metallosis, are multi factorial, including surgical technique, especially in the way preparation is made for insertion of the femoral head on the trunnion and the mdm liner into the acetabulum, as well as proper insertion of the polyethylene liner into a primary acetabular cup.Also contributing can be patient factors such as activity level and bmi, as well as implant factors." no further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text : not returned.
 
Event Description
Total hip revision.Acetabular cup, mdm liner and mdm insert.Update per medical records received: the patient was revised due to loosening of the acetabular component and metallosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSTEOLOCK BONE SCREW 24MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19090605
MDR Text Key339984715
Report Number0002249697-2024-00581
Device Sequence Number1
Product Code JDI
UDI-Device Identifier07613327035681
UDI-Public07613327035681
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K903362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number5260-5-024
Device Lot Number68627801
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2024
Initial Date FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age72 YR
Patient SexMale
Patient Weight104 KG
Patient RaceWhite
-
-