MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-170-C
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hypersensitivity/Allergic reaction (1907); Itching Sensation (1943); Urticaria (2278)
Event Date 04/09/2024
Event Type  Injury  
Manufacturer Narrative
A review of the device history record (dhr) was conducted which confirmed that the device met all quality criteria and manufacturing specifications prior to release.There is no information to indicate that a malfunction occurred.Allergic or adverse reactions are known risks of hemodialysis.The nxstage user guide and instructions for use include allergic reaction as a potential risk associated with dialysis therapy and also include warnings to monitor for potential allergic reactions.Biocompatibility of the device has been established.
 
Event Description
A report was received on 09 apr 2024 from the home therapy nurse (htn) of 69 year old male patient with a medical history of end stage renal disease, stating the patient experienced pruritus and urticaria during a hemodialysis treatment on (b)(6) 2024.Additional information was received on 09 apr 2024 from the home therapy nurse (htn) stating treatment was terminated with rinseback.Symptoms resolved following intravenous methylprednisolone (solu-medrol, dose not provided) and oral diphenhydramine (benadryl, 50mg).The patient elected to return to the dialysis facility for hemodialysis treatment later that day with no symptoms reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key19091208
MDR Text Key339985543
Report Number3003464075-2024-00076
Device Sequence Number1
Product Code KDI
UDI-Device IdentifierM535CAR170C0
UDI-Public+M535CAR170C0/$$0725401771488
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberCAR-170-C
Device Catalogue NumberCAR-170-C
Device Lot Number40177148
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/31/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient SexMale
-
-