Model Number V60
Device Problem Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/29/2024
Event Type  malfunction  
Manufacturer Narrative
Contact office phone number: (b)(4).
 
Event Description
Philips received a complaint on the v60 ventilator indicating that the device was not functioning properly.The device was reported to be outside of use at the time of the reported problem.There was no patient or user harm reported.The customer requested onsite service to be performed.A field service engineer (fse) went to the customer site and evaluated the device.The fse found that the device was producing a blower temperature high error code.The fse replaced the blower assembly and completed performance verification testing.The device passed all the included tests and was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19091251
MDR Text Key340252017
Report Number2518422-2024-18999
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/30/2024
Initial Date FDA Received04/11/2024
Date Device Manufactured12/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-