SMITHS MEDICAL ASD, INC. CADD-SOLIS AMBULATORY INFUSION PUMP; PUMP, INFUSION, PCA Back to Search Results
Model Number 2110
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Event Description
It was reported that the device had an error code 45985.Issue was not related to physical damage/abuse and there was no delay of therapy.The fault occurred during testing.There was no patient involvement and no patient harm/no adverse event reported.
 
Manufacturer Narrative
No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
One device was received for evaluation.Visual inspection found a worn dso seal, a scratched lcd lens, missing battery door, damaged battery compartment, and a bent latch lock.There was no evidence in the event history log.Functional testing was able to duplicate the reported problem.It was determined that the most probable cause is the defective pwa board.The pwa was replaced.The service history review identified no indication that the complaint was related to a service of the device within the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-SOLIS AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key19093782
MDR Text Key340681301
Report Number3012307300-2024-02492
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2110
Device Catalogue Number21-2111-0100-51
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/27/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/14/2024
Initial Date FDA Received04/11/2024
Supplement Dates Manufacturer Received04/15/2024
Supplement Dates FDA Received05/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/29/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-