MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problems Poor Quality Image (1408); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed during the device inspection, that the flex deflectable videoscope exhibited adhesive peeling of the objective lens tip and a cloudy image.There were no reports of patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and we could not specify the root cause of the peeled adhesive, however, it is probable that the defect was cause by external factors or handling.Additionally, we could not specify the root cause of the foggy image, however, it is probable that moisture entered the endoscope due to the air leak.We were unable to reproduce the reported issue.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The event can be detected by following the instructions for use: "chapter 3: preparation and inspection, item 3: inspection of the endoscope" in the instruction manual (operation section).Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19094297
MDR Text Key340803674
Report Number9610595-2024-07707
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170434662
UDI-Public04953170434662
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/03/2024
Initial Date FDA Received04/12/2024
Supplement Dates Manufacturer Received05/13/2024
Supplement Dates FDA Received05/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-