The remote service engineer (rse) spoke with the customer and confirmed the reported problem.A rse instructed biomed that there is a desaturation alarm delay time and need to go into configuration mode to check the delay time.A rse also states that if the delay is at 20 seconds, then the value must go below 85 and stay below for greater than 20 seconds.The staff stated that was too long and rse informed that it is configurable so need to work with the medical director to determine the time they would like configured for the monitors.The delay can be configured to 0 if they would like but it will generate a lot of alarms.Based on the information available and the testing conducted, the cause of the reported problem seems to be an issue with alarm management.The reported problem was confirmed.The engineer provided their analysis findings and confirmed the device was working to its specifications.There was no product malfunction, only a customer dissatisfaction and wrong perception of the device functionalities.
|