Model Number 15
Device Problem Device Sensing Problem (2917)
Patient Problem Cardiac Arrest (1762)
Event Date 03/19/2024
Event Type  malfunction  
Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.A stryker service representative performed an initial evaluation of the customer¿s device and was not able to verify or duplicate the reported issue.No accessories were returned with the device for evaluation.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
 
Event Description
The customer contacted stryker to report that paddles lead ecg on their device was not visible during intervention on a patient.In this state, the device may not be able to provide defibrillation therapy, if it were needed.This issue is patient related; however there was no adverse patient outcome reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key19095336
MDR Text Key340345953
Report Number0003015876-2024-00766
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873871713
UDI-Public00883873871713
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-001955
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2024
Initial Date Manufacturer Received 03/20/2024
Initial Date FDA Received04/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/28/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-