The customer reported the following: the anesthesiologist had the m540 docked for the start of the case and the device functioned as intended for a period of time.At some point into the procedure, that values on cockpit started alarming yellow despite alarm limits being within range.The hr value was 41 with lower and upper limits as 35 and 120 respectively, but the monitor was alarming yellow ¿hr < 50¿.When the anesthesiologist changed the alarm limit or silenced the alarm the issue remained.To attempt to fix the issue, the users tried to ¿auto set¿ all alarms, which then caused the device to not alarm when it should have.Users redocked the m540 and it seemingly resolved the issue.There was no report of any patient injury or death.
|