Complaint conclusion: as reported, the tip of a 6f/7f mynx control vascular closure device (vcd) was broken as the distal end of the sealant sleeve assembly was split apart and exposing the sealant.A new unknown mynx vcd was used for hemostasis.There were no reported injuries to the patient.The device was stored and prepped per the instructions for use (ifu).The device was not used inside of the patient but was exposed to blood during handling on the scrub table after it was opened.There were no attempts to deploy the sealant prior to the malfunction occurring.The deployer was trained.A non-sterile ¿mynx control vcd 6f/7f (ce mark)¿ was returned for investigation.Per visual analysis, the unit was thoroughly inspected observing that both button 1 and button 2 were not depressed.The syringe used during the procedure was returned attached to the unit; however, the procedural sheath was not received.The stopcock was set in the closed position, and the sealant remained in its manufactured position.The sealant sleeves presented frayed conditions that resulted in the exposure of the sealant that could be related to the reported event.Dimensional analysis was not executed as the conditions of the sleeves prevented accurate measurements for verification.Per functional analysis, the deployment mechanism of the device was completed per the intended use.During this analysis, no anomaly was observed as button 1 depression resulted in the sealant¿s deployment and button 2 depression resulted in balloon retraction to complete the expected deployed mechanism.The reported events of ¿sealant sleeves (cartridge assembly)-frayed/split/torn¿ and ¿mynx control system-deployment difficulty-premature¿ were confirmed through analysis of the returned device since the sealant sleeves were frayed and the sealant was exposed from the frayed sleeves.The exact cause of the observed conditions could not be conclusively determined during analysis.Based on the limited information available for review and product analysis, it is not possible to determine what factors may have contributed to the issue experienced.However, as the issue was reportedly found during prep of the device when the product was exposed to blood on the scrub table, prepping/handling factors are likely.It should be noted that the mynx control device is manufactured with a slit at the end of the catheter cartridge tubing.The outer sleeve is assembled with 2 side slit overlapping outer sleeves.The sealant is placed right under the outer sleeve assembly and is protected from exposing prematurely.The slits on the outer sleeve assembly are designed to decrease unsheathing force and increase deployment reliability.Refer to the diagram of the mynx control vcd within the ifu displaying the sealant sleeve with slit.If the outer sleeve is damaged/shredded during prepping phase and/or insertion into the sheath, it could cause the sealant to be exposed/swollen prematurely.As warned in the ifu, which is not intended as a mitigation, ¿do not use if components or packaging appear to be damaged or defective or if any portion of the packaging has been previously opened.¿ neither the product analysis, nor the information available for review suggest that the failures noted could be related to the design or manufacturing process of the unit.Therefore, no corrective/preventative actions will be taken at this time.
|