The diamondback 360 coronary orbital atherectomy device (oad) was used to treat a 2.5mm, 90% stenosed, heavily calcified with mild tortuosity left anterior descending artery (lad).The oad was spun for one treatment on low speed and a perforation was observed in the mid lad.The patient declined, cardiopulmonary resuscitation (cpr) was performed for twenty to thirty minutes while simultaneously performing pericardiocentesis and extracorporeal membrane oxygenation (ecmo).One liter of blood was drained from the heart.Three non-csi/abbott balloons and three non-csi/abbott stents were used and placed to successfully treat the perforation.Flow was restored and the patient was stable and placed on a ventilator and transferred to the intensive care unit (icu).The following day the patient's dependance on the ventilator was reduced.It was the physician's opinion the oad caused the perforation.
|
H6 investigation conclusion code 22: the diamondback 360® coronary orbital atherectomy system instructions for user manual states that a perforation of vessels is a potential adverse event that may occur and/or require intervention with use of the system.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|