RESPIRONICS, INC. V60 V60PLUS VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 V60PLUS VENTILATOR |
Device Problem
Tidal Volume Fluctuations (1634)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/04/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint from the customer, reporting that the v60 ventilator was not producing volumes.It was unknown how the issue was found.No patient or user harm reported.
|
|
Manufacturer Narrative
|
E1: reporting address postal: (b)(6).Reporting institution phone #: (b)(6).Reporter phone #: (b)(6).G1: contact office phone: (b)(4).
|
|
Manufacturer Narrative
|
Per the service record, a diagnosis was performed by an engineer, and it was reported that the device required a replacement blower.The repair is pending.
|
|
Manufacturer Narrative
|
H10: the service record was updated to note that the device was not in clinical use when the issue occurred.No patient or user harm reported.A service engineer (se) repaired the device, and replaced the blower assembly.The system meets the specification for the performed service and is returned to use.Although requested, the defective parts were not received at this time.A supplemental report will be submitted if the part is received and evaluated.
|
|