• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS - HOUSTON PULSE GEN MODEL 1000; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PULSE GEN MODEL 1000; GENERATOR Back to Search Results
Model Number 1000
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2023
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported that a generator was unable to be interrogated.Multiple attempts were made and different programming systems were used, but interrogation could not be completed.The generator was explanted and has not been received to date.Device history records were reviewed.The device passed all functional specifications and quality tests and was sterilized prior to distribution.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 1000
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key19096925
MDR Text Key340031475
Report Number1644487-2024-00415
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750405
UDI-Public05425025750405
Combination Product (y/n)N
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000
Device Lot Number205945
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received04/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-