RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint from the biomedical engineer (bme) on the v60 ventilator indicating that a 1122 "blower temperature high" error occurred.The device was in use on a patient at the time the reported issue was discovered; however, there was no reported harm to the patient or user.There was not any adverse outcome to patient care or therapy.The bme replaced the air inlet filter and tested the device for approximately 8 hours.The customer advised they were unable to reproduce diagnostic code 1122 - blower temperature high.No further action is required at this time.If additional information is received, the complaint file will be reopened, and a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: g - contact office phone number: (b)(4).
|
|