Manufacturer Narrative
Investigation: a terumo bct service technician checked out the machine at the customer site.A pm was completed per manufacturers instructions and the iv pole retention system was adjusted.Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No adverse event occurred and no medical intervention was required.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
 
Event Description
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.An iv pole safety collar is installed on the device.No adverse event occurred and no medical intervention was required.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
 
Manufacturer Narrative
This report is being filed to provide additional information in b.5, h.6 and h.11.Investigation: a terumo bct service technician checked out the machine at the customer site.A pm was completed per manufacturers instructions and the iv pole retention system was adjusted.One year of service history was reviewed for this device with no further issues related to the reported condition identified.The device serial number history report indicates no further related issues have been reported for this device.Correction: a field service engineer adjusted the iv pole retention system.Root cause: a root cause assessment was performed for this complaint.The root cause was determined to be related to the need for an adjustment of the iv pole retention system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key19097830
MDR Text Key340586983
Report Number1722028-2024-00141
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583810006
UDI-Public05020583810006
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3810002
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/20/2024
Initial Date FDA Received04/12/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received05/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-