G3: correction to g3 of the initial report, the aware date has been updated to march 31, 2024, based on the legal manufacturer investigation.This supplemental report is being submitted to provide additional information based on the legal manufacturer¿s evaluation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The root cause of the incident could not be determined.Based on the investigation results, it is possible that the user applied stress to the yellow connector that loosened it, and this accumulated stress caused the central part of the connector to come off.Product inspection results.Olympus has confirmed the incident.After checking the issue at olympus , we confirmed that the specifications and functionality were not met.No adverse events were reported in this complaint.Such events can be detected and prevented by following the instruction for use (ifu).Chapter 3 inspection before use.3.3 inspecting the connectors.Check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.If any irregularity is found, do not use the equipment and contact olympus.Warning do not use the equipment if any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral devices or facilities near the equipment.Olympus will continue to monitor field performance for this device.
|